Phoenix One+ Reverse Osmosis with integrated hot water sanitization

Phoenix One+ fulfills all specifications of modern water treatment systems in medical appitlications
and is designed for future standarts and developments in dialysis.
Type overview 
  Phoenix One+ Phoenix One+FH
SpecificationHot water sanitization with tankHot water sanitization with instantaneous water heater
ApplicationOne loop systems with branch line connection to the dialysis machine. Through switching on the dialysis machines in hot water sanitation hot permeate will pass from the loop in to the dialysis machine and rinse the feed line.Phoenix Multi-Loop-System
With a double ring line to the dialysis machine

The hot water sanitization is carried out independent from the dialysis machines.
Facts & Advantages
  • No storage of permeate
  • Tank is sterile ventilated in order to prevent the introduction of bacteria from the ambient air
  • Excellent energy balance, as permeate is not stored.
  • Excellent energy balance, as only the ring contents are heated
  • Drawing of heated permeate from the system is not necessary
Costs
  • No standby costs, as hot permeate is not stored
  • Adjustable tank level according to the size of the centre.
  • Very low heat loss
  • Very low power consumption
Energy consumption9 kW9 kW or 12 kW (subject to the loop length)
PipingPolyvinylidenefluoride (PDVF), bead and crevice free welding (BCF)Polyvinylidenefluoride (PDVF), bead and crevice free welding (BCF)
FrameAluminium, powder coatedAluminium, powder coated
Phoenix One+ FH

to make hot water sanitization using a energy saving instantaneous water heater

In conjunction with Phoenix Multi-Loop an efficient and energy saving hot water sanitization of the permeate system is possible.
News
Phoenix Raum Konzept im TV
Der Tag der offenen Werkstatt bei Phoenix Raum Konzept wurde vom SRF gefilmt.
Spektrum der Dialyse & Apherese
Die neuste Ausgabe Spektrum der Dialyse & Apherese erschienen. Erfahren Sie mehr...
Workshop Mikrobiologie
Nach dem Erfolg der Workshopreihe 2011 bietet Phoenix Pure Water gemeinsam mit dem Mikrobiologischen Labor Dr. Michael Lohmeyer die Veranstaltung auch in diesem Jahr wieder an...
Training ScheduleNEWSLETTERCongress CalendarFacebook